Waters Xevo TQ-S Operator's, Overview And Maintenance Manual page 52

Mass spectrometry system
Table of Contents

Advertisement

2 Preparing the Mass Spectrometer for Operation
3.
For each reservoir bottle, ensure that the end of the solvent delivery
tube is positioned so that it is close to, but does not touch, the bottom of
the bottle (see
Warning:
toxic compounds, always wear chemical-resistant, powder-free
gloves while performing this procedure.
To install the low-volume vials
1.
Remove the standard reservoir bottle, if fitted.
2.
Screw the low-volume adaptors into the manifold, and finger tighten
them.
3.
Screw the low-volume vials into the adaptors.
4.
For each low-volume vial, ensure that the end of the solvent delivery
tube is positioned so that it is close to, but does not touch, the bottom of
the vial (see
52
February 28, 2014, 715002212 Rev. C
page
53).
To avoid becoming contaminated with biohazards or
:
page
53).
Low-volume adaptor
Low-volume vial
Solvent delivery tube