Installing the new capillary
Required materials
•
Chemical-resistant, powder-free gloves
•
Needle-nose pliers
•
7-mm wrench
•
Combined 2.5-mm Allen wrench and cone extraction tool
•
Red PEEK tubing
•
LC pump
•
HPLC-grade (or better) 1:1 acetonitrile/water
•
Capillary
•
Sharp knife or PEEK tubing cutter
•
Safety goggles
Warning:
hazardous, toxic, or corrosive materials, and to avoid spreading
contamination to uncontaminated surfaces, wear clean,
chemical-resistant, powder-free gloves when working with the
probe and source components.
To install the new capillary:
1.
Use the sharp knife or PEEK tubing cutter to cut an approximately
60 cm (24 inches) piece of red PEEK tubing.
Requirement:
horizontal axis).
2.
Insert one end of the red, PEEK tubing in the probe inlet connector, and
screw the connector, finger-tight, into the PEEK union.
Replacing the IonSABRE II probe sample capillary
To avoid personal contamination with biologically
Cut the tubing squarely (that is, perpendicular to its
February 28, 2014, 715002212 Rev. C
187