4 Maintenance Procedures
5.
Slide closed the instrument's source interface door.
APPI/APCI source—replacing the APPI lamp drive seals
Warning:
APPI lamp drive assembly O-rings listed below must be renewed
at intervals not exceeding one year, exactly as described in this
section.
The following APPI lamp drive assembly O-rings must be renewed at
intervals of no greater than one year:
UV lamp bulb sealing O-ring
Mounting shaft O-rings
UV lamp mounting flange O-ring
An automatic pressure test is performed each time the source enclosure is
Tip:
closed and when the instrument starts.

Removing the APPI lamp drive assembly seals

Required materials
Chemical-resistant, powder-free gloves.
Combined 2.5-mm Allen wrench and cone extraction tool.
3-mm Allen wrench.
Small Phillips (cross-head) screwdriver.
Small Pozidriv
20-cm (8-inch) length of 4-mm nylon tube.
O-ring removal kit.
The mounting-shaft insertion tool.
A suitable, clear working area on a bench.
208
February 28, 2014, 715002212 Rev. C