Installing Ionkey Source Software - Waters Xevo TQ-S Operator's, Overview And Maintenance Manual

Mass spectrometry system
Table of Contents

Advertisement

3 Changing the Mode of Operation
µSample manager injection valve:
10. Connect the fluid inlet line to port 6 on the injection valve of the
µSample manager.
11. Connect the fluid infusion line to the grounded union connected to the
reference valve on the mass spectrometer, which provides LockSpray
for the onboard IntelliStart fluidics.
Requirement:
valve before connecting the ionKey fluid infusion line.

Installing ionKey source software

If you are installing an ionKey source on your Xevo TQ-S for the first time, you
must install the appropriate MassLynx software SCN and the ACQUITY
M-Class driver pack. For further details, see the following documents:
ACQUITY UPLC M-Class Driver Pack Installation and Configuration
Guide for detailed installation procedures, and information on using the
ACQUITY Inlet Switch Utility.
MassLynx software v4.1 and related SCN release notes for detailed
information about installing MassLynx software and SCNs.
88
February 28, 2014, 715002212 Rev. C
You must fit the supplied flow splitter kit to the fluidics
Fluid inlet line connected
to injection valve port 6