3. Basic Operation
-------- Notes ------------------------------------------------------------------------------------------------------------
Set the RMS values for the voltages of the sine wave (SIN) and the clipped sine wave (CLP)
and the Peak-to-Peak value for the voltage of the arbitrary waveform (ARB).
-----------------------------------------------------------------------------------------------------------------------------
Operation procedures for setting the DC voltage
1. Select the [DCV] item and then enter a numeric value. For how to enter values, see 3.3.5.
3.4.6
Setting the Output Voltage (Polyphase Output)
This section describes the operations in the balanced mode in a polyphase output. In the balanced
mode, the phase voltage and the line voltage are set for all phases at once. To enter the balanced
mode, select the [Osc] soft-key [6: Phase Mode] to open the phase mode set window, and select
[1:Balance] in the Mode item. For details on the unbalanced mode, see 4.15.
Operation procedures for setting the phase voltage
1. Check the icon in the output display area (see 5.1) is
system)/
and select [5:Target Phs] [4:All Phs]. Or perform the Shortcut operation
switch to [All Phs].
50
(three-phase system). If the different icon is shown, press the [Osc] soft-key
(single-phase three-wire
PROGRAMMABLE AC POWER SOURCE
+
to