Sequence - NF DP060LM Instruction Manual

Programmable ac power source
Table of Contents

Advertisement

11. Specifications

11.19 Sequence

Effective only for AC-INT, ACDC-INT, and DC-INT.
Number of memories
Number of steps
Setting range of Step Time
Intra-Step behavior
Parameter
*52 *53 *54 *55
Sequence control
*52:The output of AC phase voltage, Frequency and DC voltage of step 0 can be changed on Sequence
Edit View during output ON.
*53 : For DC-INT, the AC phase voltage, frequency, waveform, Start Phase, and Stop Phase cannot be set.
*54:The DC voltage cannot be set in the polyphase output.
*55:The Phase angle can be set only in the polyphase output. The Start Phase and Stop Phase are
specified for the L1 phase, and the component of the Phase angle setting is added to them for the
other phases.
266
5 (non-volatile)
Maximum 255 (for each sequence)
0.0010 s to 999.9999 s
Constant, Keep, Linear Sweep
Output range
AC/DC mode (The signal source is fixed to INT)
(The above 2 items are common within one sequence)
AC phase voltage, frequency, waveform
DC voltage
Start Phase
Stop Phase
Phase angle
Step termination
Jump count (1 to 9999, or infinite)
Specification of the Jump-to step
Synchronous step output (2bit)
Specification of the branch step
Trigger output
Start
Stop
Hold
Resume
Branch 1, Branch 2
All models
PROGRAMMABLE AC POWER SOURCE