Basics; Measurement Method - NF DP060LM Instruction Manual

Programmable ac power source
Table of Contents

Advertisement

It is necessary for an electric device with a large inrush current to use a power source with a
sufficient current supplying capability. With a power source that cannot supply a sufficient inrush
current, some electric devices may not start as they are not supplied with necessary power.
This product has the ability to supply peak current up to four times of the RMS rating in DP060LM /
DP120LM and three times of the RMS rating in DP180LM. You can measure the maximum inrush
current value using the current peak-hold function of the product.
4.5.2

Basics

 Output current peak-hold function holds either a positive or negative peak value, whichever is
larger in absolute value.
 Current peak-hold value is displayed in all of the Average, RMS, and Peak measurement views
in the Normal View.
 Clearing function for the current peak-hold value is provided. The clear operation clears the
current peak-hold value (all phases) to 0 Apk.
-------- Notes ------------------------------------------------------------------------------------------------------------
 The peak hold value starts to update again immediately after the clear operation. The clear
operation may not turn the current peak-hold value to 0 Apk even when no current is being
output, because of noise or other reasons.
-----------------------------------------------------------------------------------------------------------------------------
4.5.3

Measurement Method

 Operation procedure
1. Before turning on the output, clear the current peak-hold value. Following two methods are
available.
(a) Shortcut operation:
(b) Select the [Measure] soft-key  [2: Measure]. Hover the cursor over [Exec] of the [Ipk-H Clr]
item and press the ENTER key.
DP060LM/DP120LM/DP180LM
4.5 Measuring Inrush Current
131