11.27.1 Control I/O - NF DP060LM Instruction Manual

Programmable ac power source
Table of Contents

Advertisement

11.27.1 CONTROL I/O

When you enabled the external control from the menu,
You can control this product by using the external signal (or non voltage contact). The state output is
always on. The detection and state switching is done at 10 ms-cycle.
If the Remote state is achieved by the external interface, a control input is ignored.
It is not available to use 11.27.2 at the same time.
Control
External control input
input
Input level
Nondestructive maximum input
Input impedance
Control
State
Output level
output
Output impedance
Polarity
Status
Terminal
D-sub 25-pin multi-connector (rear panel, female, M2.6 screw)
*63:Low: 0.8 V or lower, High: 2.6 V or higher, the chassis potential,
*64:Sequence start and stop of the control input are effective for the Simulation as well. Also, step sync
output 1 and 2 of the control output are effective for the Simulation as well.
*65 : The memory recall input of the control input recalls the setting memory for the Normal
(Continuous), the Sequence Memory for the Sequence, and the Simulation Memory for the
Simulation. For the Sequence and Simulation, the compile data is also included.
DP060LM/DP120LM/DP180LM
Item
Output Off
Output On
Sequence start/resume
Stop of sequence
Sequence is in the Hold status
Sequence branch 1
Sequence branch 2
Memory recall (+ compile)
Memory specification 1
Memory specification 2
Clear the current peak-hold value
Power On/Off
Output On/Off*
Protection operation*
Limiter operation*
AGC/Autocal setting state*
Software busy*
Output range
Sequence operation
Step sync 1
Sequence operation
Step sync 2
Trigger
11.27 External Control Function
Description
Enabled or Disabled
TTL *63
+10 V / 5 V
Pull-up to +5 V via 47 k
Falling Off
Falling On
*64 Falling Start
*64 Falling Stop
Falling Hold
Falling Branch start
Falling Branch start
*65 Falling Recall
Specify 0 to 3
(Equivalent to memory 1 to 4,
respectively)
Falling Clear
TTL *66
220 
*67 Positive or negative
Low: Off, High: On
Low: On, High: Off (Negative)
Low: Off, High: On (Positive)
Low: Active, High: None (Negative)
Low: None, High: Active (Positive)
Low: Active, High: None (Negative)
Low: None, High: Active (Positive)
Low: On, High: Off (Negative)
Low: Off, High: On (Positive)
Low: Busy, High: Ready (Negative)
Low: Ready, High: Busy (Positive)
Low: 200 V, High: 100 V
*64
High level or Low level
*64
positive
or negative
Factory
default
negative
273

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dp120lmDp180lm

Table of Contents