SIM protocol can be tested over Ethernet using HyperTerminal.
1.
Start HyperTerminal.
2.
Enter the IP address of the pump in the Host address field. In the Port number field
enter the SIM port to be used, either 47591 or 47592. In the Connect using field,
select TCP/IP (Winsock) as shown.

Figure 72 Connect to screenshot

3.
Set the ASCII settings to the following from the File/Properties/Settings menu.

Figure 73 ASCII setup screenshot

4.
In the HyperTerminal window, type: '?T' and press the ENTER key. If successful, the
user should get a reply with the following format:
M58800880_H - Pump display terminal
Page 154