Edwards GXS Series Instruction Manual page 173

Dry screw vacuum pump
Table of Contents

Advertisement

Fault
Alarm 176.13 - No MB Inv Comms
Cause
The pump controller cannot communicate with the booster inverter during the
start‑up checks.
Remedy
Check the wiring between the pump controller and the booster inverter.
Fault
Alarm 184.10 - Dry Pump Stopped
Cause
The dry pump speed is very low. The rotor might be locked.
Remedy
Cycle the power to the pump and attempt to restart. If this fails, contact us.
Fault
Alarm 186.01 - DP INV xxxx yyyy
Cause
The dry pump inverter has raised an alarm and stopped the dry pumping system.
Remedy
Refer to
Fault
Alarm 186.01 - DP Not Running
Cause
The dry pump inverter will not start up when requested.
Remedy
Check the inverter fault history.
Fault
Alarm 186.13 - No DP Inv Comms
Cause
The pump controller cannot communicate with the dry pump inverter during
start‑up checks.
Remedy
Check the wiring between the pump controller and the dry pump inverter.
Fault
Alarm 196.01 - DP Inv xxxx yyyy
Cause
The DP 2nd dry pump inverter has raised an alarm and stopped the dry pumping
system.
Remedy
Refer to
Fault
Alarm 196.13 - No DP Inv Comms
Cause
The pump controller cannot communicate with the dry pump 2nd inverter.
Remedy
Check the wiring between the pump controller and the dry pump inverter.
Fault
Alarm 314.12 - Speed Too Low
Cause
The pump rotational speed is at least 30% less than demand for more than 3
minutes.
Remedy
Perform a mechanical check of rotation and the bearings before attempting to
restart.
M58800880_H - Fault finding
Inverter warnings and alarms
Inverter warnings and alarms
Page 169
on page 170 for more information.
on page 170 for more information.