3.7 Connect an additional RF earth (ground) (optional)
If the dry pumping system will be operated in an area subject to high Radio Frequency
(RF) emissions, in accordance with a good RF installation practise, we recommend to:
▪
use a star washer to connect the end of the earth (ground) cable connected to the
dry pumping system inlet to one of the bolts that are used to secure the inlet
flange. Refer to
skids
▪
connect an additional earth (ground) cable to the RF earth (ground) stud. A
suitable low-impedance cable must be used (for example, braided cable). Refer to
M58800880_H - Installation