Remedy
Check that cooling water is connected, switched on and is to the specification given
in
Cooling-water data
Fault
Warning 55.13/ Warning 63.13-Sensor missing
Cause
The dry pump temperature sensor may have become disconnected or failed.
Remedy
- Check that the sensor is fitted and is correctly connected.
- Check the operation of the sensor and replace it if it has failed.
Fault
Warning 71.13 - AC Sup missing
Cause
The pump controller cannot communicate with the accessory module.
Remedy
- Check that the accessory module is correctly connected to the pump and then
cycle the power to the pump.
- If this warning persists, contact us.
Fault
Warning 152.01 - Valve Not Shut
Cause
The inlet isolation valve has failed to close.
Remedy
Check the wiring and the air supply to the inlet isolation valve.*
* This warning will only clear when the valve has successfully been closed and opened.
Fault
Warning 153.01 - Valve Not Open
Cause
The inlet isolation valve has failed to open.
Remedy
Check the wiring and the air supply to the inlet isolation valve.*
* This warning will only clear when the valve has successfully been closed and opened.
Fault
Warning 176.01 - MB INV xxxx yyyy
Cause
Booster Inverter has raised a warning code.
Remedy
Refer to
Fault
Warning 176.13 - No MB Inv Comms
Cause
The pump controller cannot communicate with the booster inverter.
Remedy
Check the wiring between the pump controller and the booster inverter.
Fault
Warning 186.01 - DP INV xxxx yyyy
Cause
Dry pump Inverter has raised a warning code.
Remedy
Refer to
M58800880_H - Fault finding
on page 25.
Inverter warnings and alarms
Inverter warnings and alarms
Page 165
on page 170 for more information.
on page 170 for more information.