9.4.18 Default reset

Resets all of the non‑volatile configuration settings on the pump (for example: delays,
temperature setpoints and thresholds) to their default factory‑setting values for that
pump type.

9.4.19 Display Inverter Fault History (DP Inv Fault Hist) menu

Under the Setup menu. Scroll by pressing the UP/DOWN keys. Press ENTER to open a
sub menu or CANCEL to exit back to the Setup menu.
Table 44 Fault history menu
Fault history menu
DP Inv Fault Hist (Dry Pump Inverter Fault History)
MB Inv Fault Hist (Booster Inverter Fault History)
Each inverter fault history contains up to 32 entries, each contains one alarm and one
warning, where 1 is the most recent. Scroll by pressing the UP/DOWN keys. Press
CANCEL to exit back to the Inverter fault history menu.

9.4.20 Software version display menu

Under the Setup menu. Scroll through the software version loaded in the processors by
pressing the UP/DOWN keys. Press CANCEL to exit back to the Setup menu.
Table 45 Software display menu
Software display menu
Executive...
Support...
DP Inverter...
DP Inverter 2...
MB Inverter...
DP Inv Params...
DP Inv2 Params...
MB Inv Params...

9.4.21 Display serial number

Under the Setup menu. View the pump serial number. Press CANCEL to exit back to the
Setup menu.

9.4.22 Fit accessory menu

Under the Setup menu. Scroll by pressing the UP/DOWN keys. Press ENTER to open a
sub menu or CANCEL to exit back to the Setup menu. A security code is needed to enter
the Fit accessory menus: 538.
M58800880_H - Pump display terminal
Page 130