Chl50 - Linak Inline LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for Inline LA20:
Table of Contents

Advertisement

6. CHL50 (MEDLINE
CARELINE
®
LED indicators
LED 1 - LED 2 - LED 3
)
®
LED state
(Not listed = off)
LED 1 - 3 constantly on
LED 1 + 2 constantly on
LED 3 flashes slowly
LED 1 constantly on
LED 2 flashes slowly
LED 1 flashes slowly
LED 1 + 2 + 3 flash slowly
No light in LEDs
The CHL50 charger is an important part of the LIFT50 family . It functions as a wall
charger when mounted on the wall, but also as a more mobile charger simply placed
horizontally on its back on any surface .
Usage
• Usage temperature:
• Storage temperature:
• Relative humidity:
• Atmospheric pressure:
• Height above sea level:
• Nominal current draw:
• Power consumption (standby): < 0 .5 W
• Power consumption (charging): Max . 25 W
• Approvals:
States while charging
With battery or if no battery
mounted on charger: 90 - 100%
65 - 90%
40 - 65%
0 - 40%
Charging stopped due to low
battery temperature, high bat-
tery temperature or other error
conditions
Charging stopped due to lost
communication to battery
Page 291 of 295
+5 °C to 40 °C
-10 °C to 50 °C
20% to 80% - non-condensing
700 to 1060 hPa
Max . 3000 meters
Max . 350 mA
IEC 60601-1
ANSI/AAMI ES60601-1
CAN/CSA-22 .2 No . 60601-1