4. CB6P2 (MEDLINE
CARELINE
®
Recommendations
Battery Operation:
• If the battery voltage is at 'low level', a battery alarm beeps constantly when the HB/ACx is activated . (Low level means that battery charging is
necessary to maintain the best possible life time . Low level battery limit corresponds to approx . 19 V (+/- 5 %) .
• If the battery voltage is at 'critical level' the battery alarm function shuts down all operation immediately .
Critical level limit corresponds to approx . 17 .5 V (+/- 5 %) .
• If battery back-up is applied it only commences battery charging when it is connected to the mains .
• A battery stored at 25 °C has to be recharged every 6 - 7 months .
• Prior to first use of LINAK batteries, please make sure that they are charged for 24 hours in order to reach proper function and prolong the lifetime
of the batteries .
• The longest lifetime is obtained when the battery is fully charged .
)
®
The CB6P2 platform is introduced to obtain a powerful and optimised solution to customers looking for
existing analogue input systems . It is based on OpenBus™ technology, but to meet existing analogue
systems it has an analogue input and therefore OpenBus™ accessories cannot be connected .
Combination Overview
CB6P2 is meant for use with:
LA27 std . motor with Hall (cable type 'A')
LA27 std . motor (cable type 'B')
LA31 std . / fast motor with / without Hall
LA34 std . / small / fast motor with / without Hall (fast motor not max . load)
LA40 std . motor with / without Hall
LA44 std . / fast motor with / without Hall (fast motor not max . load)
BL1 (only with 270W transformer type) with / without Hall
HB7x, HL7x, HB8x, ACL
Page 133 of 295