Cb6 - Linak Inline LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for Inline LA20:
Table of Contents

Advertisement

3. CB6 (MEDLINE
CARELINE
®
Recommendations
• Be aware of the hand control configuration (e .g . CH1, 2, 3 should be CH1, 3, 4) .
• Same cables variants as for CB9 CARELINE and CB12 .
Connecting the system:
• Do not connect the mains cable until all actuators have been connected to the control box .
• Start by connecting the hand control to the control box . The connection at the control box is marked with "HB" .
• Connect the different actuators to the different channels on the control box . Each channel is marked with a number (e .g . "1", "2", "3" . . . .) .
• Check that all plugs are well connected and firm pushed into the connection plug . Due to the fact that LINAK control boxes are designed for a high
IP degree, a firm force can be required .
• CONNECT the mains and turn on the power!
• Finally, connect the battery (BA18) with special T-cable or normal battery cable depending on the specified system .
• The actuators can now be operated by pushing a button on the hand control . Use only one button at a time .
If the control box is equipped with special software, an initialising process might be necessary . This process is described in the software specification .
CB6 system diagram
Attention should be paid to the following:
• Control boxes must only be connected to the mains voltage specified on the label .
• The control box must be connected in such a way that the cables are not trapped, exposed to tension or sharp objects, when the application is moved
in different directions .
)
®
The control box CB6 has been specially developed for use together with LA27/LA40 actuator in
the care and rehab industry .
The control box is designed to be mounted on the actuator LA27/LA40 as with the CB9 and LA31
system .
The CB6 control box has a LED power ON indicator, detachable mains cable and strain relief for
all cables .
The control box CB6 communicates with the LA27/LA40 actuator by means of the built-in end stop
signal switches in the actuator . Due to the signal switches the power to the motor will be cut off
in the control box and not in the actuator .
Usage
• Duty cycle: 2/18; 2 minutes continuous use followed by 18 minutes not in use
• CB6 is approved according to EN60601-1 / UL60601-1
• The CB6 can only be combined with LA27/LA40 and HB30 / HB70 / HL70 / HB80 and HL80 .
Page 132 of 295