Lc1 - Linak Inline LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for Inline LA20:
Table of Contents

Advertisement

2. LC1 (MEDLINE
CARELINE
®
Top plate
up
IPX4
If the lifting column is mounted upside-down, there is NO ingress protection (IP) .
Mounting plates
Top plate
Customer production of top and/or bottom mounting plates:
LINAK can forward a 3D drawing with dimensions that comply with the customer choice of cables/plugs .
If the customer uses own top plate design, remember to make a drainage hole to drain water from the area
around the cable plug to obtain IPX4 .
)
®
Top plate
down
IPX0
The lifting column LC1 is tailor-made for use in medical applications and adds a powerful
and stable new option to the LC family of columns . The 3-stage telescopic column has
many customisation possibilities, both as to heigth and performance, and offers a more
precise choice of movement within the range of up to 4,000 N .
Usage
• Duty cycle: 10%, 2 minutes continuous use followed by 18 minutes not in use
• Operating temperature: +5 °C to +40 °C
• Storage temperature: -10 °C to +50 °C
• Compatibility: Compatible with LINAK or customer-own control boxes .
Please contact LINAK for further information .
• Relative humidity: 20% - 80%, non-condensing
• Atmospheric pressure: 700 to 1060 hPa
• Meters above sea level: Max . 3000 meters
• Approvals: IEC 60601-1
ANSI/AAMI ES60601-1
CAN/CSA-C22 .2 NO . 60601-1
Mounting guidelines
LC1 is for use in push applications to obtain IPX4 . The mounting direction must be
with the largest profile down . It is very simple to mount the LC1 in the application
using selftapping screws in the mounting holes of both endplates .
Bottom plate
Page 86 of 295

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Deskline la23Medline la23Medline la40Careline la23Careline la40Techline la23 ... Show all