4. BA21 Li-Ion (MEDLINE
®
Features and Options
• Weight:
• Housing colour:
• Protection class:
• Packaging:
• Classification:
Usage
• Compatibility:
• Duty cycle:
• Charging:
• Charging time:
• Recharging during storage: First recharge of the battery must be no later than 12 months after production date stated on the label .
• Operating temperature:
• Storage temperature:
• Relative humidity:
• Atmospheric pressure:
• Height above sea level:
• Approvals:
LED functionality:
Buzzer functionality:
The buzzer will make a warning when a button on the hand control is pressed and the battery capacity is low . The buzzer can also be activated by the
control box to signal other conditions . This must be specified in the control box software .
Mounting instructions:
The Battery Pack BA21 can be mounted in several ways on the bed/the application, either separately or together with the control box CO61 . It is
however not allowed to mount the battery in vertical position with the mounting clip pointing upwards - see illustration :
CARELINE
)
®
0 .7 kg
Light grey (RAL 7035)
IPX6 Washable DURA
TM
Every battery is packed individually and is fitted with lithium caution (transportation requirement)
Internally powered
Battery back-up for CO and CA control boxes
10%, 2 minutes continuous use followed by 18 minutes
With integrated charger in battery
Approx . 10 hours
Hereafter the battery must be recharged at least every 12 months .
+5 °C to +40 °C
-10 °C to +40 °C (+10 °C to +25 °C recommended)
The batteries must be stored in an applicable storage room without direct sunlight .
20% to 80% - non-condensing
700 to 1060 hPa (3000 m)
Max . 3000 meters
IEC60601-1, ANSI/AAMI ES60601-1, CAN/CSA-22 .2 No 60601-1, IEC62133, UL2054,
UN38 .3 (needed for transport of lithium batteries)
The BA21 Li-Ion back-up battery pack has been specially developed for
use with the new control boxes COxx and CAxx, e .g . CO61 and CA40,
etc . It is a low weight battery with built-in charger and high performance
and safety .
LED
Indication of operation
Solid yellow
Charging
No LED light
Fully charged
Flashing yellow
Error during charging
Page 212 of 295