Chl40 - Linak Inline LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for Inline LA20:
Table of Contents

Advertisement

3. CHL40 (MEDLINE
CARELINE
®
Usage
• Usage temperature:
• Storage temperature:
• Relative humidity:
• Atmospheric pressure:
• Height above sea level:
• Nominal current draw:
• Power consumption (standby): Max . 2 .5 W
• Power consumption (charging): Max . 19 W
• Approvals: (pending)
LED functionality
The charger indicates whether it is connected to mains (green LED) or whether the battery is being charged (orange LED) .
Charger mounting/position
The LIFT40 charger CHL40 can be placed for use in two different ways . It is designed for mounting flat on the wall by using minimum 2 of 3 screws in
the Ø4 mm holes in the charger body . Screw types and wall plugs may vary depending on wall material .
The charger can also be placed flat on a table or desk . The dedicated rubber studs on the charger back ensure that the charger stays safely at the
same position during charging (see drawings) .
Front View
When mounting CHL40 on wall,
please use minimum two of the 3
dedicated mounting holes placed in
body of charger.
Mounting information
CHL40 is mounted by means of minimum 2 screws (not supplied by LINAK) .
Screw type: The screw type depends on the wall type and has to be defined by the service technician .
Washer type: ISO7089 / M5, d1 = 5 .3 mm / d2 = 10 mm / s = 1mm .
The CHL40 charger must be mounted with minimum two of the three screws possible .
The mounting screws for the control box and the charger must be tightened with a maximum torque of 1 Nm .
To be able to comply with the IPX4 rating, the CHL40 must hang on the wall .
The charger must be disconnected from mains in cleaning situations .
Recommendations
• Only use original LINAK components and accessories for full compatibility .
• Only use CHL40 charger for BAL40 battery charging .
• The charger CHL40 is specifically designed to charge the BAL40 battery .
• Special care should be taken when mounting the CHL40 .
• If the CHL40 is mounted correctly, the CHL40 complies to IPX4 .
• If the CHL40 is mounted incorrectly, water may enter the screw holes resulting in IPX4 non-compliance and cause malfunction and hazardous situations .
• In cleaning situations, the charger must be disconnected from mains .
)
®
+5 °C to 40 °C
-10 °C to 50 °C
20% to 80% - non-condensing
700 to 1060 hPa
Max . 3000 meters
Max . 500 mA (CH01 spec)
IEC 60601-1, ANSI/AAMI ES60601-1, CAN/CSA-22 .2 No . 60601-1, Australian deviation, Canadian deviation
Mounting holes
The CHL40 charger is an important part of the LIFT40 family . It functions
as a wall charger when mounted on the wall, but also as a more mobile
charger simply placed horizontally on its back on any surface .
When placing CHL40 on table of
shelf, 4 rubber studs on the back
adds stability and friction to the
charger / battery combination.
Studs for table placement
Page 278 of 295
Back View
Studs for table placement

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Deskline la23Medline la23Medline la40Careline la23Careline la40Techline la23 ... Show all

Table of Contents