18. HD80 (MEDLINE
CARELINE
®
Standard HD80 - HD84C1J0550004-200120012D1C000
Item number J90208
This hand control can be used as a combination of a hand control and the ACO . It has two levels of operation, where the first is a patient mode with
regular operations like hi/lo and trend/anti-trend . Use the magnet key to operate the next level, care mode, where it is possible to lock functions .
The LEDs show which functions are locked and which are not .
Magnet key - article no . 0858008
Warning
• Do not sit or lie on the hand control . It can cause unintended movement of the bed .
• Inform the customer that after loss of mains power, the lock state is reset to the default setting . Be aware of a special setup for a magnet lock of
low power system in case of power down on mains . Also be aware that the lock is reset when running on battery or when powered down .
• Inform the customer that using the magnet key cannot wake up a low power system or a system running on battery . The system will wake up when
activating a key and then the magnet key can unlock the system .
• Inform the customer that a powerful magnetic field may change the lock state .
• Always use O-rings on connectors and cable locks .
• There is a risk that items with internal magnet for mounting instead of hook can disturb function of cardiac pacemaker, implantable cardioverter
defibrillators or magnetic implants! .
Recommendations
• Inform the customer to use only the magnet key supplied by LINAK . We also recommended to make a functional test of the application before putting it
into operation .
• Clean the hand control regularly to ensure good hygiene standards .
• When replacing a defective HD80, check that the new HD80 has exactly the same specification and functionality .
• Do not submerge the hand control under water .
• Unless otherwise specified or agreed by LINAK, the hand control is only intended to be used on LINAK systems .
• When changing hand controls for OpenBus™, the power must be switched off .
• It is recommended to check the hand control and cable for damage and holes made by violent handling before washing the bed or at least once a year .
• In order to maintain the flexibility of the cables, it is important that a coiled cable is placed in such a way that the cable's own weight does not strain the
coil during the washing process .
For hand controls with magnets:
• If hand controls with magnets are hooked on a smooth surface, a movement or twisting of the cable, for example during transport, can cause the hand
control to move and result in damage if the cable gets squeezed somewhere .
• The force of the magnet depends on the thickness of the lacquering, the lacquering type, stickers, steel thickness etc . It is the responsibility of the
customer to verify that the holding force on the application is acceptable .
• It is the responsibility of the user/operator to evaluate any possible risk caused by use of magnets .
• It is recommended to have a parking place for the hand control on the application where the customer ensures that the hand control does not fall off .
)
®
The HD80 makes it possible to have two hand controls in one unit . The hand control
is equipped with a magnet locking function, making it possible to have two levels of
operation – one for the patient and relatives and one for the caregiver staff . The HD80
provides a great overview using LED indication of functions being locked or unlocked .
The hand control is designed to work with OpenBus™ systems .
Usage
• Usage temperature:
• Storage temperature:
• Compatibility:
• Relative humidity:
• Atmospheric pressure:
• Height above sea level: Max . 3000 meters
• Approvals:
5 ºC to 40 ºC
-10 ºC to +50 ºC
Compatible with CB6 and OpenBus™ control boxes .
Please contact LINAK
20% to 80% - non-condensing
700 to 1060 hPa (3000 m)
IEC60601-1, ANSI/AAMI ES60601-1 and
CAN/CSA-22 .2 No 60601-1