7. LA28 Compact ( MEDLINE
Recommendations
• LINAK control boxes are designed so that they will short-circuit the motor terminals of the actuator(s), when the actuator(s) are not running .
This solution gives the actuator(s) a higher self-locking ability . If the actuator(s) are not connected to a LINAK control box, the terminals of the motor
must be short-circuited to enable selflocking of the actuator .
• Min . stroke length for LA28 with splines is 80 mm
• The current supply to LINAK actuators must be cut off in case of overload and when the actuators reach end position .
• Ambient operating temperature is 22 °C .
• LA28 Compact cannot be used in pull applications, unless fitted with an aluminium back fixture .
Further information:
Noise level:
• LA28: dB(A) 45; measuring method DS/EN ISO 3743-1, actuator not loaded
• LA28S: dB(A) 54; measuring method DS/EN ISO 3743-1, actuator not loaded
Material:
• The piston rod eyes are "crimped" in
place and cannot be screwed loose .
8. LA29 (MEDLINE
CARELINE
®
CARELINE
TECHLINE
)
®
®
®
The LA28 Compact is a small and powerful actuator designed for use in system solutions
for healthcare equipment or industrial applications . Ideal applications are for example
wheelchairs, treatment chairs, patient lifts or beds .
Usage
• Duty cycle: Max 10 % or max . 2 min . continuous use followed by 18 min . not in use .
• Ambient temperatures: + 5 °C to + 40 °C
• Compatibility: Compatible with LINAK control boxes . Please contact LINAK .
• Approvals: IEC 60601-1, ANSI/AAMI ES60601-1 and CAN/CSA-22 .2 No 60601-1
"Crimped" piston rod eye
)
®
Usage
• Duty cycle: 10% or 2 minutes continuous use followed by 18 minutes not in use
• Usage temperature: +5 oC to +40 oC
• Storage temperature: -10 oC to 50 oC
• Compatibility: Compatible with LINAK control boxes . Please contact LINAK .
• Relative humidity: 20% to 80% - non-condensing
• Atmospheric pressure: 700 to 1060 hPa
• Height above sea level Max . 3000 meters
• A CE Declaration of Conformity has been issued
This particular LINAK
actuator is the ideal choice in medical equipment where power
®
and speed are required, but where space is limited .
Page 56 of 295
Protective
cap
Piston rod without eye (are not tested)