17. HB400 (MEDLINE
CARELINE
®
Magnet key - article no. 0858008
Recommendations
• It is recommended to make a functional test of the LINAK system in order to ensure that control box and hand control are communicating correctly .
• Unless otherwise specified or agreed with LINAK, the hand control is only intended to be used for LINAK systems .
• Inform the customer only to use the magnet key supplied by LINAK .
• Do not submerge the hand control into water .
• It is recommended to check the hand control and cable for damage and holes caused by violent handling before washing the application or at least
once a year .
• In order to maintain the cable flexibility, it is important to place a coiled cable in such a way that its own weight does not strain the coil during the
washing process .
• When changing hand controls for OpenBus™ systems, the power must be switched off .
• When a defective HB400 is replaced, check that the new HB400 has exactly the same specification and functionality .
Warnings
• Do not sit or lie on the hand control . It can cause unintended movement of the application .
• The application manufacturer must write an end-user manual based on the LINAK user manual which also includes relevant warnings,
information on how to carry out regular inspection and a functionality description . End users must be trained in all functions .
• Inform the customer that using the magnet key cannot wake up a low-power system or a system running on battery .
The system will wake up when a key is activated and the magnet key will then unlock the system .
• Inform the customer that a powerful magnetic field may change the locking state .
• Always use O-ring on connectors and cable locks .
)
®
The HB400 hand control is designed for a wide range of applications such as hospital
beds, nursing home beds and other medical applications such as treatment chairs and
couches .
Usage
• Operation temperature: +5 °C to + 40 °C
• Storage temperature: -10 °C to + 50 °C
• Relative humidity: 20% to 80% - non-condensing
• Atmospheric pressure: 700 to 1060 hPa
• Height above sea level: Max . 3000 meters
• Flammability rating: HB
• Latex free: Yes
• Approvals: IEC 60601-1
Page 199 of 295
ANSI/AAMI ES60601-1
CAN/CSA-22 .2 No 60601-1