Mjb5 Plus Ubl (Int) - Linak Inline LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for Inline LA20:
Table of Contents

Advertisement

12. MJB5 Plus UBL (int) (MEDLINE
Usage
Compatibility:
Operation temperature:
Storage temperature:
Relative humidity:
Atmospheric pressure:
Operational meters above sea level:
Latex free:
Approvals:
Not available in version with internal UBL
Standard functionality:
UBL: The UBL LED can be switched on and off via the patient control (Key 1) or theattendant control (Key 2) .
External Switch (S1/S2): The external switch is connected to S1: Pin 2 & 8/S2: Pin 2 and 9 on Port 2, Port 4 or Port 5 .
The Switch input functionality can be enabled/disabled via the attendant control . The enable/disable status (switch status) is indicated on the OpenBus .
The switch input functionality is as standard to be used with a NO switch . When the switch is activated (NC), a notification is sent on the OpenBus (switch
notification) .
CARELINE
)
®
®
The modular junction box MJB5 Plus is designed for use together with OpenBus™
control boxes .
The MJB5 Plus makes it possible to connect multiple hand controls and attendant
controls . It can even be used for charging or to connect the Under Bed Light and 3
party products .
MJB5 Plus with Under Bed Light (int.) and Switch Input (S1/S2)
MJB5 Plus versions 505-010 and 505-020:
The MJB5 Plus is a modular junction box with 3 different options, Under Bed Light
(UBL) and 2 different switch inputs, S1 and S2, which can be used for an external
switch .
UBL: The MJB5 Plus with UBL gives a guiding light when the patient leaves the bed and
makes it easy to find the way back to bed at night without disturbing other patients .
The MJB5 Plus has an LED integrated in the MJB5 Plus housing which makes it easy to
use .
External Switch (S1/S2): It is possible for the customer to connect a switch directly
to the MJB5 Plus . This can be used with a customised switch or control .
All OpenBus products
+5
o
C to +40
o
C
-10
C to + 50
C
o
o
20% to 80% non-condensing
700 to 1060 hPa
Max . 3000 meters
Yes
IEC60601-1, ANSI/AAMI ES60601-1, CAN/CSA-22 .2 No 60601-1,
Port 3
Page 236 of 295
Port 5
OpenBus™ connection /
S1 pin 2 & 8 / S2 pin 2 & 9
Port 4
OpenBus connection /
S1 pin 2 & 8 / S2 pin 2 & 9
Port 2
OpenBus connection /
S1 pin 2 & 8 / S2 pin 2 & 9
Port 1
OpenBus connection
rd

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Deskline la23Medline la23Medline la40Careline la23Careline la40Techline la23 ... Show all

Table of Contents