13. MJB5 Plus UBL (ext) (MEDLINE
Usage
Compatibility:
Operation temperature:
Storage temperature:
Relative humidity:
Atmospheric pressure:
Operational meters above sea level:
Latex free:
Approvals:
Port 3 is not available
Standard functionality:
UBL: The UBL LED can be switched on and off via the patient control (Key1) or the attendant control (Key2) .
External Switch (S1/S2): The external switch is connected to S1: Pin 2 & 8/S2: Pin 2 and 9 on Port 2, Port 4 or Port 5 .
The switch input functionality can be enabled/disabled via the attendant control . The enable/disable status (switch status) is indicated on the OpenBus .
The switch input functionality is as standard to be used with a NO switch . When the switch is activated (NC), a notification is sent on the OpenBus (switch
notification) .
CARELINE
)
®
®
The modular junction box MJB5 Plus is designed for use together with OpenBus™
control boxes .
The MJB5 Plus makes it possible to connect multiple hand controls and attendant
controls . It can even be used for charging or to connect the Under Bed Light and 3
party products .
MJB5 Plus with Under Bed Light (ext.) and Switch Input
MJB5 Plus versions 506-010 and 506-020:
The MJB5 Plus is a modular junction box with 3 different options, Under Bed Light
(UBL) and two switch inputs, S1 and S2, which can be used for an external switch .
UBL: The MJB5 Plus with UBL is a simple solution to prevent fall accidents and make
the patient feel safe . The UBL makes it easy to find the way back to the bed at night
without disturbing other patients .
The MJB5 Plus has an external LED cable (0964135) which is connected to Port 5 .
The external LED cable makes it more flexible to use the UBL . It can be moved from
side to side of the bed or can be placed at the foot of the bed .
External Switch (S1/S2): It is possible for the customer to connect a switch directly
to the MJB5 Plus . It can be used with a customised switch or control .
All OpenBus products
+5
C to +40
C
o
o
-10
C to + 50
C
o
o
20% to 80% non-condensing
700 to 1060 hPa
Max . 3000 meters
Yes
IEC60601-1, ANSI/AAMI ES60601-1, CAN/CSA-22 .2 No 60601-1,
Port 5
Port 4
Port 2
Port 1
Page 242 of 295
OpenBus™ connection / Ext . UBL cable / S1 pin pin 2 & 8 / S2 pin 2 & 9
OpenBus connection / S1 pin 2 & 8 / S2 pin 2 & 9
OpenBus connection / S1 pin 2 & 8 / S2 pin 2 & 9
OpenBus connection