Lc3 - Linak Inline LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for Inline LA20:
Table of Contents

Advertisement

3. LC3 (MEDLINE
CARELINE
®
LC3 mounting guidelines
LC3 is for use in push or pull applications and can be mounted in both directions – largest profile down or largest profile up .
LC3 must be specified for specific mounting direction .
Mounting direction according to item number nomenclature .
Observe no
drain holes
Observe
drain holes
IPX6 is only available when LC3 is mounted with the largest profile up .
IPX6 is only applicable when the LC3 column is fully retracted .
Note: The cable outlet for motor connection can be positioned at the top (smallest profile) or from the side of the column .
)
®
Top plate up
The LC3 2-stage and 3-stage set the standard for vertical lifting columns to high-end
medical applications . This compact lifting column fulfils the market requirements for a
solid and stable lifting column .
Usage:
• Duty cycle: 10%, 2 minutes continuous use followed by 18 minutes not in use
• Operation temperature:
• Storage temperature:
• Compatibility:
• Relative humidity:
• Atmospheric pressure:
• Meters above sea level:
• Approvals:
Top plate down
IPX4
Page 90 of 295
+5 °C to + 40 °C
-40 °C to + 70 °C
Compatible with LINAK control boxes .
Please contact LINAK .
20% to 80% - non-condensing
700 to 1060 hPa
Max . 3000 meters
IEC 60601-1
IEC 60601-1-6
ANSI/AAMI ES60601-1
CAN/CSA-C22 .2 No . 60601-1
Observe no
drain holes
Observe no
drain holes
IPX4
IPX6