12. HB70 (MEDLINE
CARELINE
®
Usage
• Compatible with most LINAK control boxes .
• Approved according to: EN 60601-1, EN 60335-1 and UL 60601-1 as part of a LINAK actuator system
Recommendation
• It is not possible to combine HB7x with the binary based CB9 . .PM/PN .
• The IPX6 Washable version has a special adhesive for the front covers .
• The HB75xE0 used together with CB140 will give trend and anti-trend on channel 1 and 2 of the control box when using the last button row .
• All front covers use the codes W0 (not Washable) and WW (Washable) Memory:
Memory:
• The memory and parallel functions require the control box to have a microprocessor .
• When storing a memory position on the control box, the actuators must run to the desired position and the "store" button (S) must be pushed .
Then the desired memory position button (1, 2 or 3) must be activated within 2 seconds .
13. HB80 (MEDLINE
CARELINE
®
Usage
• Usage temperature: 5 ºC to 40 ºC
• Storage temperature: -10 ºC to +50 ºC
• Compatibility: Compatible with many LINAK control boxes . For further questions, please ask your local LINAK .
• Approvals: IEC60601-1, ANSI/AAMI ES60601-1 and CAN/CSA-22 .2 No 60601-1
The HB86 version has a shorter distance between the buttons and cannot be approved according to EN IEC60601-2-52 Application Environment
4 for care beds used in Domestic area (or EN1970) . HB80 is designed and tested in accordance with EN60601-2-52 cl . 201 .11 .6 .6 .101 (machine
washable medical beds) .
The HB80 must hang vertically from it's hook during the washing process .
In order to maintain the flexibility of the cables, it is important that a coiled cable is placed in such a way that the cable's own weight does not
strain the coil during the washing process .
Recommendations
• Clean the hand control regularly to ensure good hygiene standards .
• When a defective HB80 is replaced, check that the new HB80 has exactly the same specification and functionality .
• Do not submerge the hand control under water .
• Unless otherwise specified or agreed by LINAK - the hand control is only intended to be used on LINAK systems .
• When changing hand controls for OpenBus™ systems, the power must be switched off .
• It is recommended to check the hand control and cable for damage and holes made by violent handling before washing the application or at least
once a year .
• It is recommended to have a parking place for the hand control on the application, where the customer ensures that the hand control does not fall off .
For hand controls with magnets:
• If hand controls with magnet are attached to a smooth surface, a movement or twisting of the cable, for example during transport, can cause the
hand control to move and result in damage if the cable is squeezed .
• The force of the magnet depends on the thickness of the lacquering, the lacquering type, stickers, steel thickness etc . It is the responsibility of the
customer to verify that the holding force on the application is acceptable .
• It is the responsibility of the user/operator to evaluate any possible risk caused by use of permanent magnets .
Warnings
• Do not sit or lie on the hand control . It can cause unintended movement of the application .
• There is a risk that items with internal magnet for mounting instead of hook can disturb function of cardiac pacemaker, implantable cardioverter
defibrillators or magnetic implants!
)
®
)
®
The HB80 hand control has an optimised ergonomic design shaped for the hand . The hand control is suitable for
all kinds of MEDLINE and CARELINE applications such as hospital beds, patient lifts, treatment and examination
couches etc .
The HB80 hand control is available in versions with up to 10 or 12 activation buttons .
The HB70 offers simultaneous drive of multiple actuators which can be
used for the memory options . The hand control HB70 can be used for both
OpenBus™ and analogue systems and comes in 3 colours: black, dark
grey and light grey .