Ca30/Ca40 - Linak Inline LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for Inline LA20:
Table of Contents

Advertisement

1. CA30/CA40 (MEDLINE
®
LED indicator
CA30-CA40 is equipped with a three-colour LED for indication of mains or battery operation .
Connected to MAINS
LED colour
Indication of operation
On mains, not activated by hand or foot control.
Green
The system is working ok and is ready for normal operation
Yellow
On mains, not activated by hand or foot control.
The system is defective and should not be operated .
Yellow
On mains, activated by hand or foot control.
The system is working .
Not connected to mains but with BATTERY back-up
LED colour
Indication of operation
Orange
On battery, activated by hand or foot control.
The system is working
No LED
On battery, not activated by hand or foot control.
or CA30-CA40 not connected to mains
CARELINE
)
®
Page 117 of 295
The control boxes CA30 and CA40 are developed
as part of a new control box platform for the
care and rehab industry . The control boxes can
be mounted separately on the application by
means of unique slide-on brackets, but also on
several actuator models using a specially designed
actuator bracket .
Features and options
• Duty cycle:
10 % - 2/18 min . on/off continuos use .
Maximum power is 120 W for 80 seconds
and 60 W for 40 seconds at 25 °C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Deskline la23Medline la23Medline la40Careline la23Careline la40Techline la23 ... Show all