Mjb5 Plus Aoc - Linak Inline LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for Inline LA20:
Table of Contents

Advertisement

14. MJB5 Plus AOC (MEDLINE
Usage
• Compatibility: All OpenBus control boxes except CB20 .
• Operation temperature: +5 C to +40 C
• Storage temperature: -10 C to + 50 C
• Relative humidity: 20% to 80% non-condensing
• Atmospheric pressure: 700 to 1060 hPa
• Operational meters above sea level: Max . 3000 meters
• Latex free: Yes
• Approvals: IEC60601-1, IEC60601-1-6
ANSI/AAMI ES60601-1
CAN/CSA-22 .2 No 60601-1
MJB5 Plus AOC Usage
Compatibility: All OpenBus control boxes except CB20 .
Port 3 is not available
The product label is added for correct connection to the ports:
The MJB5 AOC has 4 ports .
Port 1 is for OpenBus connection .
Port 2 has 4 common ground switch inputs or an OpenBus hand control .
Port 4 has 8 common positive switch inputs or an analogue hand control .
Port 5 has 8 common positive switch inputs or an analogue hand control .
Port 1 for connection to the control box and port 2 can be used for an OpenBus control . The standard inputs are made so that a system containing
an analogue hand control can be used for an OpenBus control box in port 4 or port 5 .
CARELINE
)
®
®
Port 5
Port 4
Port 2
Port 1
The modular junction box MJB5 Plus is designed for use together with OpenBus™
control boxes .
The MJB5 Plus makes it possible to connect multiple hand controls and attendant
controls . It can even be used for charging or to connect the Under Bed Light and 3
party products .
MJB5 Plus – Analogue to Openbus Converter (AOC)
MJB5 Plus versions 509
The AOC is an Analogue to OpenBus™ Converter . It is the perfect match for especially
medical applications where the customer needs to add own products or 3
controls in an OpenBus system, e .g . multi-purpose foot switches .
The MJB5 AOC is available with or without power request .
Common positive
(input 9, 10, 11, 12, 13, 14, 15 or 16)
Common positive
(input 1, 2, 3, 4, 5, 6, 7, 8)
OpenBus connection or 4 inputs with common GND
(input 17, 18, 19 or 20)
OpenBus connection
Page 248 of 295
rd
party
rd

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Deskline la23Medline la23Medline la40Careline la23Careline la40Techline la23 ... Show all