19. HD80 JUMBO (MEDLINE
Warnings
• Do not sit or lie on the hand control . It can cause unintended movement of the application .
• Always use O-ring on connectors and cable locks .
• There is a risk that items with internal magnet for mounting instead of hook can disturb cardiac pacemaker functions, implantable cardioverters,
defibrillators or magnetic implants .
Recommendations
• Clean the hand control regularly to ensure good hygiene standards .
• When a defective HD80 is replaced, check that the new HD80 has exactly the same specification and functionality .
• Do not submerge the hand control under water .
• Unless otherwise specified or agreed by LINAK, the hand control is only intended to be used on LINAK systems .
• When changing hand controls for OpenBus™, the power must be switched off .
• It is recommended to check the hand control and cable for damage and holes made by violent handling before washing the bed or at least once a year .
• In order to maintain the flexibility of the cables, it is important that a coiled cable is placed in such a way that the cable's own weight does not strain
the coil during the washing process .
Hand controls with magnets:
• If hand controls with magnet are hooked on a smooth surface, a movement or twisting of the cable, for instance during transport, can cause the hand
control to move and result in damage, if the cable is squeezed somewhere .
• The force of the magnet depends on the lacquering thickness, the lacquering type, stickers, steel thickness etc . The customer has the responsibility
to verify that the holding force on the application is acceptable .
• The user/operator is responsible for evaluating any potential risk caused by the use of magnets .
• It is recommended to have a parking spot for the hand control on the application where the customer ensures that the hand control does not fall off .
20. HL70 (MEDLINE
CARELINE
®
Recommendations
• To switch between locked and unlocked position a small knob between the two push buttons has to be turned 20° by use of a special key .
The key is for the use of the nursing staff only, there are two types, one is made of plastic the other metal .
• For all types: Attention should be given to ensure that the channels shown correspond to the channels available on the chosen control box .
• The HL70 must hang vertically from its hook during the washing process . In order to maintain the flexibility of the cables, it is important that
a coiled cable is placed in such a way that the cable weight does not strain the coil during the washing process .
CARELINE
)
®
®
The HD80 JUMBO is a hand control with an optimised ergonomic design and functions that
are activated via dome buttons .
Usage
• Usage temperature:
• Storage temperature:
• Compatibility:
• Relative humidity:
• Atmospheric pressure: 700 to 1060 hPa (3000 m)
• Height above sea level: Max . 3000 meters
• Flammability rating:
• Approvals:
)
®
5º C to 40º C
-10º C to +50º C
Only compatible with CBJ Care
20% to 80% - non-condensing
UL94-V2
IEC60601-1, ANSI/AAMI ES60601-1 and
CAN/CSA-22 .2 No 60601-1
The HL70 is a hand control with integrated locking function, where a selec-
tive locking of the different functions is available by use of a special key .
The HL70 is an alternative to the HB70 combined with an attendant control
panel such as the ACL .
Usage
• Exchangeable with HB70
• Operation temperature: +5 °C to +40 °C
• Storage temperature: -10 °C to +50 °C
• Relative humidity: 20% to 80% - non-condensing
• Atmospheric pressure: Max . 3000 meters
• Compatible with many LINAK control boxes
• Approvals: IEC 60601-1
IEC 60601-1-6
ANSI/AAMI ES60601-1
CSA CAN/CSA-C22 .2 NO . 60601-1