5. Information on specific control boxes

Please be aware if the control box is not visible after mounting, all information regarding limitation of use shall be marked on the end product .
Output voltage
On control boxes connected to the mains the voltage of the actuator output is dependent on load, and the no-load voltage can reach 50 V .
Control boxes connected to a battery can reach a voltage of 30 V during charging and no load .
For all control boxes with battery
Prior to first use of LINAK batteries, please make sure that they are being charged 24 hours in order to reach proper function and prolong the lifetime of
the batteries .
Warning
Please observe the following maintenance, replacement, and disposal requirements to ensure a safe and reliable operation .
Maintenance of batteries
The batteries are to be replaced after 4 years at the latest . Perhaps earlier, dependent on the pattern of use . Frequent and high-powered discharges reduce
the battery life . For an optimum lifetime the product must be connected to the mains voltage as often as possible . It is recommended that the batteries
are to be charged at least every 3
function at least once every year .
Replacement of batteries
The batteries must only be replaced by the same type of batteries or mechanical and electrical equivalent types .The batteries must be new or maintained by
means of charging at least every 3
codes may lead to a severely reduced life time expectancy .
Before mounting ensure that the battery set is correctly connected, compare with the drawing in the battery room, and check that no connectors are loose .
Warnings
From the factory the battery room is hermatically separated from the electronics room . When replacing the batteries this separation must not be
damaged or modified as this may allow penetration of battery gas into the electronics room with risk of explosion .
When replacing batteries in waterproof products (IPX5 and IPX6) precautions must be taken that the sealing material (silicone ring or joint filler)
is not damaged and that it is correctly placed in the groove . Hereafter the screws in the cover are to be fastened with appox . 1 Nm . If the seal is
damaged it must be replaced by a new silicone string (LINAK article no . 0008004 for a roll of 100 metres) .
Disposal
The batteries, which are lead-acid batteries, can be returned to LINAK or disposed in the same way as car batteries .
Warnings
The battery room is supplied with ventilation that ensures correct and necessary airing of the battery room . This airing must not be blocked or
covered as a positive pressure may occur with risk of explosion .
If the product has been exposed to mechanical overload (lost on the floor, collision/squeezing in the application or a powerful stroke) the product
must be sent to an authorised workshop for control of the hermetic separation between the battery and electronics rooms .
Using control boxes with speed control
The common way of carrying out a speed control of the actuators is by using PWM, switching the motor on/off at a high frequency .
In rare cases, while switching the motor on/off, we have experienced that the coupling between the actuators and application frame and ground is too
high, thus generating a slight current noise which exceeds the allowable EMC limits . The current in question is extremely low and is in no way related to
any personal or patient risk .
The coupling is defined by the mechanical layout of the application, and no real guidelines can be given . Using plastic bushings or similar can improve the
application .
If an application faces this issue, it can easily by solved by connecting ground of the CB box to the application frame-, through a cable with built in serial
connection of a resistor and a capacitor .
The EMC test defined in IEC60601-1-2, applicable for all medical products, will show if this is an issue concerning the specific applications .
If you need more information or have any issues on this subject, please contact your LINAK A/S Sales contact .
Design criteria when using a customized CB
When using a customized CB together with a LINAK actuator, the interface connection between the CB vs . Actuator as well as the purpose of the actuator
(with or without feedback) must be considered .
Actuators with feedback
E .g . potentiometer solutions have certain technological characteristics due to the design that might cause quality issues if used outside the specification
range .
The connection interface
The connection interface is not only the compatibility of the plug types used, e .g . male mini-fit to female mini-fit plug type .
It is also considering the contact transition, i .e . the cable as well as the cable connector, the connecting plug, the material surface of the plug PIN, the
soldering of the PCB plug connector etc .
When however using a LINAK Actuator with feedback AND having a bad connection interface towards a Customized CB – LINAK only
guarantees the feedback voltage to be within 500 mV (+/- 250 mV).
rd
month - otherwise will the batteries have reduced capacity due to self-discharge . It is recommended to test the battery
month . The batteries, which make a set, must be supplied with identical production codes . Mismatching of production