4. Information on specific columns

1. BL1 (MEDLINE
CARELINE
®
Input/output specifications: dual Hall, digital positioning
Column with Dual Hall (BL141H) are equipped with two hall sensors, A and B and a spindle magnet . In this way you can have pulses from the column
when it moves .
The feedback system has a 8P magnet which gives 16 shifts in pulses per spindle turn .
Hall Output:
4 mm spindle = 0,25 mm per pulse and 400 pulses per 100 mm stroke .
9 mm spindle = 0,5625 mm per pulse and 178 pulses per 100 mm stroke .
Important:
The motor must always be short-circuited to obtain self-locking in accordance with the rated label value .
TECHLINE
)
®
®
The BL1 is a 3-part lifting column designed to be used for example in hospital beds,
nursing home beds, treatment chairs, couches and dental chairs .
The lifting column is compact and has a long stroke length . The 3-part guidance enables an
overlap between the individual profiles, which ensures a high degree of stability .
The lifting column has an open spindle actuator with a chain drive inside which is
practically noiseless .
Usage
• Duty cycle: 10%, 2 minutes continuous use followed by 18 minutes not in use
• Usage temperature: + 5 °C to + 40 °C
• Storage temperature: Max 50 °C
• Approvals: IEC60601-1, ANSI/AAMI ES60601-1,
CAN/CSA-22 .2 No 60601-1
IEC 60601-1-6
IEC 60601-1-2