14. HB100 (MEDLINE
CARELINE
®
Usage
• Operation temperature: +5 °C to + 40 °C
• Storage temperature:
-10 °C to + 50 °C
• Relative humidity:
20% to 80% - non-condensing
• Atmospheric pressure: 700 to 1060 hPa
• Height above sea level: Max . 3000 meters
• Approvals:
IEC 60601-1:2005 + Amd .1:2012 (Consolidated version IEC 60601-1:2012 Ed . 3 .1)
IEC 60601-1-2:2014 Ed . 4
• Compatibility:
All OpenBus
• Flammability rating:
V2
• Latex free:
Yes
Replacing the cable
The cable for the HB100 can be replaced if damaged . To remove the cable, the cable lock must first be unlocked . This is done by moving the lock-pin
clockwise with a screwdriver or another small object, until a red marker shows . When inserting a new cable, the lock pin must be moved counter-
clockwise to secure a fastened cable connection .
Recommendations
• Keep the hand control upright when washing
• Do not submerge the hand control in water
• Unless otherwise specified or agreed with LINAK, the hand control is only intended to be used for LINAK systems
• When changing hand controls for OpenBus™ systems, the power must be switched off
• It is recommended to check the hand control and cable for damage and holes caused by violent handling before washing the application or at least
once a year .
Warnings
• Do not sit or lie on the hand control . It can cause unintended movement of the application
• The application manufacturer must write an end-user manual based on the LINAK user manual which also includes relevant warnings,
information on how to carry out regular inspection and a functionality description . End-users must be trained in all functions .
)
®
control boxes
TM
Page 194 of 295
The HB100 is an intelligent hand control with the LINAK
Solution . It features a 2 .4" colour display with a full navigation menu
keypad, thus allowing the display to be dynamic without physically
having to change the hand control . The HB100 will automatically scan
the system and only show the features available to create a user-friendly
experience .
The HB100 will be available with different software versions:
• HB110: The LINAK Weighing Solution . Together with the QLCI2, the
HB110 is capable of displaying the weight of a patient as well as
setting the Out Of Bed feature .