Mjb2 - Linak Inline LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for Inline LA20:
Table of Contents

Advertisement

8. MJB2 (MEDLINE
CARELINE
®
Usage
• Usage temperature:
+5 °C to +40 °C
• Storage temperature:
-10 °C to +50 °C
• Compatibility:
Connection to LINAK OpenBus and analogue control boxes
• Relative humidity:
20% to 80% – not condensing
• Atmospheric pressure:
700 to 1060 hPa
• Latex free:
Yes
• Approvals:
IEC 60601-1
ANSI/AAMI ES60601-1
CAN/CSA-22 .2 No 60601-1
Recommendations
• Always use locking mechanism and O-ring .
• Unused socket(s) must be fitted with blind plug(s) to ensure the IP degree .
• When mounting, a screw torque 0 .8-0 .9 Nm is recommended .
• Hot-plugging: removing or adding any OpenBus™ cables is not allowed when the control box is powered by mains supply .
If still required, follow this procedure:
1 . Remove mains and wait for 5 seconds
2 . Connect or disconnect the required cables
Non-observance of this procedure may lead to a damaged OpenBus driver curcuit .
The risk of a damaged circuit increases, if the accessory shows a high starting current or inrush current .
Warnings
• Using a wrong type of MJB2, for instance 10 wires in an OpenBus system, can lead to unintended movement or no movement .
• Using wrong screws or the wrong torque can lead to cracks in the housing .
• The cable is not to be exposed to high pull force or sideway traction .
)
®
The MJB2 is a compact 2-port repeater designed for use together
with analogue or OpenBus™ control boxes . It is optimised for use
in systems where 1 extra port is needed for easy connection of a
hand control, a foot switch or an accessory like the UBL . It is easy to
integrate in a wide range of healthcare applications such as hospital
beds, surgery tables, and treatment chairs .