10. MJB5 Plus SMPS (MEDLINE
Usage
• Compatibility:
• Operation temperature:
• Storage temperature:
• Relative humidity:
• Atmospheric pressure:
• Operational meters above sea level: Max . 3000 meters
• Latex free:
• Approvals:
Standard functionality:
Port 1 is to be connected to the OpenBus control box .
The electronic device must be connected to PORT 2 . The power on port 2 can be switched
on/off via the patient control (key 1) or the attendant control (key 2), the status is
indicated on the OpenBus . Switching on/off the power supply is useful, for instance when
the SMPS is used together with a bedside lamp . If an error occurs, this is indicated on the
OpenBus . As standard, it is switched off .
Modular plug cable, narrow/wide alignment
grooves:
Modular plug with wide alignment groove.
To be used with LINAK products .
Can be connected to all ports in the MJB5 Plus,
both ports with narrow and wide alignment
grooves .
CARELINE
)
®
®
The modular junction box MJB5 Plus is designed for use together with OpenBus™
control boxes .
The MJB5 Plus makes it possible to connect multiple hand controls and attendant
controls . It can even be used for charging or to connect the Under Bed Light and 3
party products .
MJB5 Plus with Switch Mode Power Supply (SMPS)
5V SMPS, version 502-010.
The SMPS is to be used where there is a need for power supply near the bed . For
example, to charge electronical devices .
It is also possible to connect a bedside lamp from our 3rd party supplier . This will be
connected directly to the MJB5 Plus port 2 via a modular plug . The SMPS indicates
with an LED on the hand control whether the power supply is switched on or off .
All OpenBus products
+5
C to +40
C
o
o
-10
C to + 50
C
o
o
20% to 80% non-condensing
700 to 1060 hPa
Yes
IEC60601-1, IEC60601-1-6
ANSI/AAMI ES60601-1
CAN/CSA-22 .2 No 60601-1
Page 224 of 295
+5V Power supply
-Pin 2 (GND)
-Pin 7 (reserved)
-Pin 8 (reserved)
Port 2
-Pin10 (output power)
Port 1
OpenBus™ connection
Modular plug with narrow
alignment groove.
To be used with 3
party products .
rd
Can only be connected to ports in the
MJB5 Plus with narrow slit . This is to
prevent 3
party products to interfere
rd
with the OpenBus™ connections .