5. BAL50 (MEDLINE
CARELINE
®
Usage
• Compatibility:
• Duty cycle:
• Charging:
• Charging state:
• Recharging during storage:
• Operating temperature:
• Charging temperature:
• Charging time:
• Storage temperature:
• Relative humidity:
• Atmospheric pressure:
• Meters above sea level:
• Approvals:
Mounting
Mounting of the BAL50 battery will follow the placement of the COL50 and CHL50 .
See user manual for further information in the respective product sections .
Deep discharge protection
The BAL50 has a deep discharge protection to extend the battery life . The deep discharge protection is activated when the battery is discharged .
If the battery is completely discharged, the charging will be started at a very small rate to protect the battery . Depending on the battery state, it may
take several hours to get to the normal charging state .
For long-term storage, remove the battery from the application and store separately .
Transportation
The lithium ion batteries must be packed and transported according to applicable regulations . Always ask your local transportation provider how to
handle the transportation of lithium ion batteries .
Recommendations
• Charge the battery fully before first use .
• Adhere to the battery storage temperature or else the product life and performance will be reduced .
• Allow the battery to settle to room temperature before use or charging .
• Only use correct LINAK
charger (CHL50 or integrated charger in COL50) .
®
• Do not exceed the duty cycle as it will shorten the product life and reduce performance .
• BAL50 is intended for use in indoor applications, however not in indoor pool environments .
• Recharge the battery before storage if it has been completely discharged .
)
®
COL50 and CHL50
Max . 10% (or 2 min . continuous use followed by 18 min . without use) at max . discharge current (10 A)
Via external wall charger CHL50 or via COL50 with integrated charger
Maximum 30% when shipped from LINAK
First recharge of the battery must be no later than 12 months after production date stated on the label .
Hereafter the battery must be recharged at least every 12 months .
+5 °C to +40 °C
+5 °C to +40 °C
Charging at elevated temperatures can impact the charging time
Approx . 4 hours
+10 °C to +40 °C (+10 °C to +25 °C recommended)
The batteries must be stored in an applicable storage room without direct sunlight .
20% - 80% – non-condensing
700 to 1060 hPa
Max . 3000 meters
IEC 60601-1
ANSI/AAMI ES60601-1
CSA CAN/CSA-C22 .2 NO . 60601-1
IEC 60601-2
IEC62133-2
The BAL50 Li-Ion battery pack is a part of the LIFT50 product
series specially developed for patient lifts .
LIFT50 is a complete system consisting of the control box
COL50, a battery BAL50 and an external charger CHL50 in a
flexible solution .