Acl; Aco - Linak Inline LA20 User Manual

Linear actuators and electronics
Hide thumbs Also See for Inline LA20:
Table of Contents

Advertisement

Recommendations
• The customer responsibility includes making a proper design of the cable strain relief inside the side rail panel .
• The customer should consider the existence of vibrations when defining and specifying the housing, i .e . we recommend the customer to carry out a
vibration test on the final product .
• The customer must ensure a proper IP rating/test
• The customer must ensure proper drop testing according to IEC60601-2-52 §201 .15 .3 .4 .1 . In this clause there is an additional reference to
IEC60601-2-31 .
• The customer is responsible for correct mounting of the PCBA . Among other things, it means
- ensuring proper and safe mounting of the PCBA into e .g . the side rail
- ensuring proper and correct mounting between key pad connection tails and the ACK PCBA
- ensuring proper and correct mounting of the key pad
- the customer should consider proper precautions against ESD (Electrostatic discharge) .
• When handling ESDS (Electrostatic Discharge Sensitive) devices – e .g . during transport, storage, handling, production or mounting in an application
- exposure to harmful ESD must be avoided .
• Consider proper creepage and clearance measures to fulfil IEC 60601 . With One MOPP (One Means Of Patient Protection / Secondary side of the
actuator system)
• It is not recommended to dismount the membrane front cover after mounting as this may cause damage .
4. ACL (MEDLINE
CARELINE
®
5. ACO (MEDLINE
CARELINE
®
Recommendations
• Always use Locking ring and cables with O-rings .
• Locking ring and cables with O-rings must be fitted to ensure IP degree .
• If other front covers than standards are requested, the front cover guidelines should be consulted .
N.B.
• Cables are inclusive an O-ring .
)
®
The ACL (Attendant Control Lock) box is a one turn button box for various applications where the patient
positioning must be carefully controlled by the medical staff .
The ACL disconnects all functions on the hand control either by means of turn button or turn key .
)
®
The Attendant Control OpenBus
be used as hand control keys or lock-outs . The lock-out function can be made visable by using yellow LEDs .
Usage
• Operation temperature: +5 °C to +40 °C
• Storage temperature:
• Relative humidity:
• Atmospheric pressure:
• Flammability rating:
• Approvals:
In order to comply with the norm, the ACO must hang vertically from its hook during the washing process .
(ACO) is a cost optimised and compact unit with up to 21 buttons that can
TM
-10°C to +50 °C
20% to 80% non-condensing
700 to 1060 hPa (3000 m)
V2
IEC 60601-1
IEC 60601-1-6
ANSI/AAMI ES60601-1
CAN/CSA-C22 .2 NO . 60601-1
Page 184 of 295