13. CO53 (MEDLINE
CARELINE
®
Thermal protection
CO53 is thermally protected . The CO53 protects itself by shutting down . Once the thermal condition is normalised, the CO53 will recover and allow
further operation .
Recommendations
NOTE - HOT PLUGGING
Removing or adding any OpenBus™ cables are not allowed when the control box is powered by mains supply!
If needed anyway, follow the below procedure:
1 . Remove mains and wait 5 sec .
2 . Mount or dismount the required cables
If this procedure is NOT followed, it may result in a damaged OpenBus driver circuit . The risk of a damaged circuit increases if the accessory has a high
start current (in rush current) .
NOTE - Use of internal mains signal in software or on OpenBus
Please be aware when using the internal mains signal on control boxes with SMPS that the mains signal may take up to 6 seconds before disappearing
after mains has been removed .
Battery operation:
BATTERY LEVELS
Battery High: > 20 V
Battery Medium: 18-20 V
Battery Low: < 18 V
• If the battery voltage is at 'Medium level', a battery alarm beeps as long as a key is activated . (Medium level means that battery charging is necessary
to maintain the best possible lifetime) .
• If the battery voltage is at 'Low level', the battery alarm function shuts down all movement immediately . The OpenBus is still active for approx . 15
seconds . If trying to operate the system anyway, the battery could get deep drained or the actuator system could get damaged . When at low battery
level (which is critical), there is a risk that the processor will incorrectly monitor the end of stroke . Crashing the actuator could be a result .
• The CO53 with battery back-up only starts battery charging when it is connected to the mains .
)
®
The LINAK
190 Watt power supply, a small and compact design and the ability to handle multiple hand controls
or other OpenBus accessories for applications with a need for more advanced features .
Features and options
• Control concept: Relays (EAS protection)
Usage
• Operation temperature: +5 °C to +40 °C
• Storage temperature: -10 °C to +50 °C
• Relative humidity: 20% - 80% not condensing
• Atmospheric pressure: 700 to 1060 hPa
• Meters above sea level: Max . 3000 meters
• Approvals:
- normal
- alarm
- critical
control box CO53 is a new member of the OpenBus™ range . The control box offers a
®
IEC60601-1
ANSI/AAMI ES60601-1
CSA CAN/CSA-C22 .2 NO . 60601-1