Assembly

This section describes the procedures to assemble U-LLG150/U-LLG150 and power cords.
A
B
C
3
D
E
6
1

Attaching U-LLG150/U-LLG300

Hold the incidence side connector A of U-LLG150/U-LLG300 with
1
hand, and insert it completely to the end of the liquid light guide
insertion port of U-LGPS.
Loosen the clamping screws B C of U-LLGAD using the Allen
2
screwdriver provided with the microscope.
If the clamping screws B C are not fully loosened, the tip
TIP
of the emission side connector D contacts the clamping
screws B C and the emission side connector may not be
inserted to the correct position.
Insert the emission side connector D of U-LLG150/U-LLG300
3
completely to the end of the liquid light guide insertion port of
U-LLGAD.
Tighten the clamping screws B firmly using the Allen screwdriver
4
provided with the microscope.
Pull the emission side connector D toward you to make sure
TIP
that it does not come out.
Tighten the clamping screws C firmly using the Allen screwdriver
5
provided with the microscope.
If the clamping screws B C are not tightened firmly,
WARNING
U-LLG150/U-LLG300 may come off from U-LLGAD and a
fire may be caused.
6
Attach U-LLGAD to the illuminator.
7
Tighten the clamping screw E of the illuminator using the Allen
screwdriver provided with the microscope.
Do not turn ON the lamp while U-LGPS is not attached to
WARNING
the microscope using U-LLG150/U-LLG300 and U-LLGAD.
The light from this lamp contains the UV light. Looking
directly at the light may damage your eyes. Also, it could
cause a fire.
U-LLG150/U-LLG300 and power cords are vulnerable to
NOTE
bending and twisting. Be careful not to apply excessive
force.