To Output Arbitrary Waveform

1. Press the Wave Gen button on the front panel to enable the AWG function and enter
the WAVEFORM function menu.
2. Connect the T3DSO2000 to a PC which is running the EasyWave software using
a USB cable.
3. Double click the EasyWave icon to open the program.
4. Select a waveform file or draw an arbitrary waveform; then click the Send wave option
which at the bottom of the interface and it will pop-up the dialogue which shows as the
picture below:
5. Click the Store location area and select the location to keep the arbitrary waveform.
6. Click the Send option, then the oscilloscope will pop-out the message "Dealing with
waveform data, please wait..."
7. A few seconds later the oscilloscope will prompt "Succeed to save waveform", and it
will enter the WAVEFORM function menu, the Wave Type sets to ArbX (the location
chose above)
8. To delete the arbitrary waveform, press the Delete softkey.
9. Press the Setting softkey to enter the SETTING function menu.
10. Press the Output Load softkey to select High-Z or 50Ω.