To Enable Quick-Cal

Quick-Cal can automatically calibrate the T3DSO2000 when ambient temperature
variations are affecting the measurement accuracy. If the temperature environment that
the T3DSO2000 is working in is not stable, perform the following steps:
1. Press the Utility button on the front panel to enter the UTILITY function menu.
2. Press the Next Page softkey to go to the second page of the UTILITY function menu.
3. Press the Quick-Cal to select On.
If the Quick-Cal has been set to On before powering on, and the current vertical scale is
2mV or 5mV of any channel, the oscilloscope will perform the quick calibration procedure
when it powers on. During the calibration, most of the buttons are disabled.
Also, when the Quick-Cal is On and the current vertical scale is 2mV or 5mV of any
channel, the oscilloscope will perform the quick calibration procedure if the operating
temperature changes by more than 2 degrees.