AWG Self-Cal
If the ambient operating temperature changes more than 5°and the instrument runs more
than more 30 minutes, run the AWG self cal routine to reduce the environmental thermal
change effects
1. Press the Wave Gen button on the front panel to enter the WAVEFORM function
menu.
2. Press the Setting softkey to enter the SETTING function menu.
3. Press the AWG Self Cal softkey to perform AWG self-calibration program.
Figure 106: AWG Self Cal
4. When the self-calibration program is completed, press the Run/Stop button on the
front panel to exit the interface according to the information on the screen.