Trigger Level - Teledyne T3DSO2000 Series User Manual

Hide thumbs Also See for T3DSO2000 Series:
Table of Contents

Advertisement

Trigger Level

Trigger level and slope define the trigger point,
Positive Slope
Negative Slope
Trigger Level
Trigger Point Trigger Point
Input Signal
The trigger level can be adjusted for a selected analog channel by turning the Trigger
Level Knob.
Push the Trigger Level Knob to set the level to the middle of the waveform. If AC
coupling is used, pushing the Trigger Level knob sets the trigger level to approximately
0 V.
The position of the trigger level for the analog channel is indicated by the trigger level icon
(If the analog channel is on) at the left side of the display. The value of the analog
channel trigger level is displayed in the upper-right corner of the display.