To Adjust The Vertical Scale; To Adjust The Vertical Position - Teledyne T3DSO2000 Series User Manual

Hide thumbs Also See for T3DSO2000 Series:
Table of Contents

Advertisement

To Adjust the Vertical Scale

The vertical scale can be adjusted in Coarse or Fine mode.
Coarse adjustment (turning counterclockwise as an example): Set the vertical scale
in 1-2-5 step namely 1mv/div, 2 mV/div, 5 mV/div, 10 mV/div ...10 V/div.
Fine adjustment: Further adjust the vertical scale within a relatively smaller range to
improve vertical resolution. For example: 2 V/div, 1.98V/div, 1.96V/div, 1.94 V/div ...1
V/div.
If the amplitude of the input waveform is slightly greater than the full scale under the
current scale and the amplitude would be to low if the next scale is used, fine adjustment
can be used to improve the amplitude of waveform display to view signal details or
perform vertical based measurements.
Press the CH1 button on the front panel; then press the Adjust softkey to select the
desired mode. Turn the VERTICAL Scale Knob to adjust the vertical scale (clockwise to
reduce the scale and counterclockwise to increase).
The scale information in the channel label at the right side of the screen will change
accordingly during the adjustment. The adjustable range of the vertical scale is related to
the probe ratio currently set. By default, the probe attenuation factor is 1X and the
adjustable range of the vertical scale is from 1mv/div to 10 V/div.
Note: Push the VERTICAL Scale Knob to quickly switch between Coarse and Fine
adjustments.

To Adjust the Vertical Position

Turn the VERTICAL Position Knob to adjust the vertical position of the channel
waveform. Turn the knob clockwise to increase the vertical position and the channel
waveform moves up while counterclockwise to reduce the vertical position and the
waveform moves down. Push the knob to set the vertical position of the channel waveform
to zero.
During the adjustment, the vertical position information Volts Pos displays at the bottom of
the screen. The table below shows the range of vertical position according to the volt
scale.
1.800.561.8187
Volt Scale
1 mV/div ~ 100 mV/div
102 mV/div ~ 1 V/div
1.02 V/div ~ 10 V/div
www.
Range of Vertical Position
±1V
±10 V
±100 V
information@itm.com