To Adjust REF Waveform Display

1. Please refer to the 'To Display REF Waveform' above to display the desired
reference waveform.
2. Use the Vertical Scale knob and Vertical Position knob for Math/Ref to adjust the
vertical scale and position of the reference waveform. The vertical scale and position
information displays in the middle of the screen.
The initial values displayed in the middle of the screen are the setups when the reference
waveform was saved.
Note: The vertical Variable knob and the vertical Position knob only work for the

To Clear REF Waveform Display

T3DSO2000 does not have the 'Clear' option under the REF WAVE function menu. To
clear the appointed reference waveform, you can save a new reference waveform to the
same location to replace it.
1.800.561.8187
120
Figure 65: Reference Waveform
reference waveform and the math waveform. Please note the difference with
the channel vertical system.