Radio Frequency Interface (Rfi) Board - GE Vivid 3 Pro Service Manual

Hide thumbs Also See for Vivid 3 Pro:
Table of Contents

Advertisement

GE M
S
EDICAL
YSTEMS
D
2300164-100, R
IRECTION
5-4-5

Radio Frequency Interface (RFI) Board

5-4-5-1
General Description
The Radio Frequency Interface (RFI) board has been introduced in the manufacture of Vivid™ 4
systems, as a result of certain hardware components having become obsolete.
The functionality of the following boards is implemented on the RFI board:
Front-End Controller (FEC) board
RF & Tissue (RFT) board
Image Port (IMP) board
NOTE:
On Vivid™ 4 systems with the RFT configuration, the FEC, RFT and IMP board functionality is still
implemented on the BEP via separate boards. These are described in this chapter in the following
sections:
on page 5-24;
Tasks Performed by the RFI board:
System Master Clock generator
Transmitter Control
ATGC Control
Receive Focusing Control
Analog Test Signal Generator
High Voltage Control and supervision
Probe Sensing
Control of MUX board, Front Board, and Beam Former
Temperature Sensing for TEE probes
Step Motor Control for TEE probes
5-4-5-2
Signal Flow
The digitized signals from the Beam Formers are connected to the RFI module. The Digital Signal
Processor on the RFI board, together with software on the BEP, performs the adequate signal
conditioning for the different data types; Tissue, Doppler and Color Flow.
In addition, Doppler Audio Processing is performed by the RFI board. (The Doppler Spectrum
Processing is done by the BEP).
The RFI extracts Tissue Data, Spectrum Data and IQ & Color Data from the digital data stream on its
input, and it also performs RF filtering and different types of Tissue Processing.
5-4-5-3
RF Processing
Different types of RF processing is performed, depending on later usage of the data:
RF Demodulation
Digital Time Gain Compensation
Filtering
Decimation
5-20
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
7
EVISION
Front End Controller Board (FEC) (RFT)
Image Port Board (IMP)
Section 5-1 - Overview
V
IVID
on page 5-23;
RF and Tissue Processor Board (RFT)
on page 5-25.