The test set's signal input is subject to the limits and characteristics contained in
C A UTI ON
Table 44
To prevent damage to the test set's hardware components, do not apply the input
signal to the test set's signal input connector until the input attenuator (Option 001)
has been set by the phase noise software, which occurs at the connection diagram.
Table 44 Test set signal input limits and characteristics
Limits
Frequency
Maximum Signal Input Power
At Attenuator Output, Operating Level Range:
Downconverters:
Characteristics
Input Impedance
AM Noise
Refer to
interconnections.
276
on page 276.
RF Phase Detectors
Microwave Phase Detectors
Internal AM Detector
Agilent N5502A/70422A
Agilent N5507A/70427A
Figure 210
on page 277 for more information about system
50 kHz to 1.6 GHz (Std.)
50 kHz to 26.5 GHz (Option 001)
50 kHz to 26.5 GHz (Option 201)
Sum of the reference and signal input power
shall not exceed +23 dBm
0 to +23 dBm (Signal Input)
+15 to +23 dBm (Reference Input)
0 to +5 dBm (Signal Input)
+7 to +10 dBm (Reference Input)
0 to +20 dBm
0 to +30 dBm
+5 to +15 dBm
50 Ω nominal
DC coupled to 50 Ω load
Agilent E5505A User's Guide