It is normal to hear clicking when the Auto Alignment function is on. See
NOTE
"Running Internal Alignments" on page 18

Why Aren't All the Personality Options Loaded in Memory?

Many measurement personality options are available for use with this instrument.
If the option is loaded in the instrument, you must also have a license key entered,
to use it.
Some versions of instrument hardware may not have enough memory to
accommodate all the options that you have ordered. If this is the case you will need
to swap the applications in and out of memory, as needed. It may also be possible
to upgrade your hardware to have more memory. Contact your local sales or
service office.

If Option 117 is Loaded in Memory

If you ordered Option 117, the Secure Memory Erase option was loaded into the
memory of your instrument and licensed at the factory but not enabled. DO NOT
enable this option until you have fully configured your instrument (set the
Config I/O
configured the power-up state you desire). After the security mode is enabled, the
main Flash memory becomes read-only, rather than read/write memory. This
makes it impossible to add or remove personalities or their license keys, or
otherwise change the instrument's configuration, unless you do a
Security, Secure Erase All
inoperable. You will then need to run the upgrade procedure (on the Agilent Web
site) to reinstall the firmware and completely reconfigure your instrument. Refer to
"Managing Security" in Chapter 4 of the "User's and Programmer's Reference,
Volume 1."
When you are satisfied with the configuration of your PSA analyzer, enable the
Secure Memory Erase option by pressing
The security mode will then be enabled the next time the analyzer's power is
cycled.
NOTE
When Option 117 is present, the number of personalities that can be loaded into
memory is limited to 3 or 4.
Chapter 1
calibrated measurement. To meet its specifications, the analyzer must meet
operating temperature conditions.
settings as needed, added or removed options and licenses, and
Turning on the Analyzer for the First Time
for more information.
procedure. This procedure will render the instrument
System, Security, Security, Enabled
Installation and Setup