4.9.6 O
Sensor calibration not overcome
2
WARNING !! Risk of DM failure
The oxygen sensor calibration is not overcome when the detected voltage value is
not included between 43 mV and 67 mV.
WARNING !! Malfunctions risk - calibration failed
•
•
•
CAUTION - Replacing the oxygen cell
The oxygen cell must be replaced when, at the end of the calibration phase, appears
a detected voltage value not included between 43 mV and 67 mV and/or if the
system displays the relevant alarm message.
To order the replace sensor and to dispose of the worn one, please see
chapter 7 : Maintenance.
O
Morpheus_ND
Check if the oxygen sensor is installed and electrically connected to the
anaesthesia unit.
Check if the oxygen sensor is worn out (the oxygen detection cell is worn out);
replace the oxygen sensor.
Repeat the oxygen sensor calibration.
Sensor calibration failed: please see chapter 6 : Alarms and Troubleshooting.
2
oxygen sensor calibration
The
procedure has not been successfully
completed.
O
Sensor calibration failed.
2
4-35