6.1.3 A3 - ALARMS limits : gas sensor

Touching this icon, a window showing the available Alarm limits appears and the User can possibly
set the Min and Max values of each single alarm
6-8
WARNING !! Risk for Patient / User injury
The anaesthesia unit used in the same health environments can have different
preset configurations of alarm limits.
Verify that the preset alarm limits are appropriate for the new patient and adjust
the alarm limits on values suitable to the new condition of use.
The alarms limits setting could make the alarm not properly working.
Select: gas sensor area
The Alarm Limits screen appears.
Select
YES: to quit Alarm page; the alarm set
will NOT be saved.
NO: it remains in Alarm page.
Select
YES: to quit Alarm page; the alarm set
will be saved.
NO: it remains in Alarm page.