5.8.2 APCV-TV

(Volume Targeted ) Pressure controlled ventilation,
synchronised with patient's breathing and with guaranteed
current volume.
The system displays all PRP relative to the operative
mode set.
APCV-TV
is a pressure controlled ventilation, synchronised with the patient's breathing (automatic
Pinsp) with guaranteed current volume (Vte).
The system generates a ventilation at automatic inspiration pressure (automatic Pinsp), in order for
the expired volume to equal the volume set (Vte).
During the inspiratory phase, the Unit generates an automatic flow ( Slope ). When the pressure
reaches the control value inside the airway (automatic Pinsp, at maximum PMax), this pressure level
is kept constant by the Unit until the end of the inspiration that you can set using the (RR) and the I:E
ratio.
Use the settable parameters to define an inspiratory trigger (Tr. I) used to set a flow expressed in
litres per minute (or a pressure in cmH
spontaneous breathing attempt.
5-32
O) that represents the limit for detecting the patient's
2
User manual, DU33ND101